using Aitex.Core.RT.Routine;
using Aitex.Core.RT.SCCore;
using Venus_RT.Devices;
using MECF.Framework.Common.Routine;
using System.Diagnostics;
using System.Collections.Generic;
using Venus_Core;
namespace Venus_RT.Modules.PMs
{
    class GasBoxLeakCheckRoutine : PMRoutineBase, IRoutine
    {
        private enum LeakCheckStep
        {
            kPumpToBasePressure,
            kPumpingDelay,
            kLeakCheckDelay,
            kEnd,
        }
        public double LeakRate { get; private set; }

        private readonly double _GasFlow = 1.0;

        private int _basePressure = 100;
        private int _leakcheckPumpTime = 180;
        private int _leakcheckHoldTime = 300;

        private double _startPressure = 0;
        private double _endPressure = 0;
        private double _leakRate = 30.0;
        private double _leakCheckBasePressure = 1;
        private List<int> _gasLineNums = new List<int>();

        Stopwatch _leakCheckTimer = new Stopwatch();


        public GasBoxLeakCheckRoutine(JetPM chamber) : base(chamber)
        {
            Name = "GasBox Leakcheck";
        }

        public RState Start(params object[] objs)
        {
            if (CheckLidLoadLock() &&
                CheckSlitDoor() &&
                CheckTurboPump())
            {
                Reset();

                _basePressure = SC.GetValue<int>($"{Module}.Pump.PumpBasePressure");
                _leakcheckPumpTime = SC.GetValue<int>($"{Module}.Pump.LeakCheckPumpingTime");
                _leakcheckHoldTime = SC.GetValue<int>($"{Module}.Pump.LeakCheckHoldTime");
                _leakRate = SC.GetValue<double>($"{Module}.Pump.LeakRate");
                _leakCheckBasePressure = SC.GetValue<double>($"{Module}.Pump.LeakCheckBasePressure");

                // Extract line numbers which need do Leakcheck from config file
                _gasLineNums.Clear();
                var lineNums = SC.GetStringValue($"{Module}.Pump.LeakCheckGasLineNums").Split(',');
                int nNum;
                foreach(string num in lineNums)
                {
                    if(int.TryParse(num, out nNum))
                    {
                        if(nNum > 0 && nNum <= 8 && !_gasLineNums.Contains(nNum))
                        {
                            _gasLineNums.Add(nNum);
                        }
                    }
                }

                if(_gasLineNums.Count == 0)
                {
                    Stop($"No Gasline need do LeakCheck, please check the config item{Module}.Pump.LeakCheckGasLineNums");
                    return RState.Failed;
                }


                PreSetValves();

                return Runner.Start(Module, Name);
            }
            return RState.Failed;
        }

        public RState Monitor()
        {
            Runner.Wait((int)LeakCheckStep.kPumpToBasePressure, ()=> { return _chamber.ChamberPressure <= _basePressure; })
                .Run((int)LeakCheckStep.kPumpingDelay,          LeakCheckPumping,                                   PumpingDelay)
                .Run((int)LeakCheckStep.kLeakCheckDelay,        StartLeakCheck,                                     _leakcheckHoldTime * 1000)
                .End((int)LeakCheckStep.kEnd,                   CalcLeakCheckResult,                                _delay_50ms);

            return Runner.Status;
        }

        public void Abort()
        {
            CloseAllValves();
        }

        private bool LeakCheckPumping()
        {
            foreach(var num in _gasLineNums)
            {
                _chamber.FlowGas(num, _GasFlow);
            }
            _chamber.OpenValve(ValveType.PV11, true);
            _chamber.OpenValve(ValveType.PV21, true);
            _chamber.OpenValve(ValveType.PV31, true);
            _chamber.OpenValve(ValveType.PV41, true);

            _leakCheckTimer.Restart();
            return true;
        }

        private bool PumpingDelay()
        {
            if (_leakCheckTimer.ElapsedMilliseconds >= _leakcheckPumpTime * 1000)
            {
                if (_chamber.ChamberPressure <= _leakCheckBasePressure)
                    return true;
                else
                {
                    Runner.Stop($"GasBox Leakcheck失败, 腔体压力 [{_chamber.ChamberPressure}]mTor, 高于LeakCheck Base Pressure: [{_leakCheckBasePressure}] mTor");
                }
            }

            return false;
        }

        private bool StartLeakCheck()
        {
            _startPressure = _chamber.ChamberPressure;

            Notify($"PM 压力开始值 {_startPressure} mt");

            _chamber.TurnPendulumValve(false);
            return true;
        }

        private bool CalcLeakCheckResult()
        {
            _endPressure = _chamber.ChamberPressure;

            LeakRate = (_endPressure - _startPressure) * 60.0 / _leakcheckHoldTime;

            if (LeakRate < _leakRate)
            {
                Notify($"GasBox Leakcheck完成, 压力结束值: {_startPressure} mt, 漏率:{LeakRate} mt/min");
            }
            else
            {
                Stop($"GasBox Leakcheck失败, 腔体漏率 [{LeakRate}] mt/min, 高于 [{_leakRate}] mt/min");
            }

            _chamber.StopAllGases();
            _chamber.TurnPendulumValve(true);

            _chamber.OpenValve(ValveType.PV11, false);
            _chamber.OpenValve(ValveType.PV21, false);
            _chamber.OpenValve(ValveType.PV31, false);
            _chamber.OpenValve(ValveType.PV41, false);

            return true;
        }

        private void PreSetValves()
        {
            _chamber.OpenValve(ValveType.FastPump, false);
            _chamber.OpenValve(ValveType.SoftPump, false);
            _chamber.OpenValve(ValveType.TurboPumpPumping, true);
            _chamber.OpenValve(ValveType.Guage, true);
            _chamber.OpenValve(ValveType.GasFinal, true);

            _chamber.TurnPendulumValve(true);

            _chamber.StopAllGases();
            _chamber.OpenValve(ValveType.PV11, false);
            _chamber.OpenValve(ValveType.PV21, false);
            _chamber.OpenValve(ValveType.PV31, false);
            _chamber.OpenValve(ValveType.PV41, false);
        }
    }
}